FGF REFOLDING KINETICS MEASUREMENTS USING APPLIED PHOTOPHYSICS SX 20
STOPPED-FLOW
Dr. Jihun Lee
8/24/07
Sample preparation
1. Prepare
samples according to the “Refolding kinetics” formula (below)
(if Cm is higher
than 2.5M, denature protein with 3.0M)
à 20 uL
x 12 shots x sample #
(e.g. if Cm=1.0
M, prepare buffer from 0.25 M GuHCl to 0.9M GuHCl)
2.
Equilibrate the denatured protein overnight
3.
Measurement
4. Usually
repeat the full data collection twice to get the averaged rate constants
Click here to download the excel
spreadsheet (see below)

To turn on the instrument
1. Turn on
the “LAMP” switch
2. Ignite
the lamp (push the red button)
3. Record
the lamp hour in the log-in book
4. Open N2
gas à adjust the pressure (use the adjuster behind “sample
handling unit”)
1:10 mixing (0.25 mL:2.5 mL) : 30 psi
5. Turn on
the “SYSTEM” switch
6. Turn on
the computer à log-in “SX users”
7. Open “Pro-Data
SX” software
8. Set the
working directory
1) Click
“Pro-Data View” icon in the SX window
2) Specify
the directory for data saving and analysis
(C:\Documents and Settings\SX Users\DATA
USERS\...)
3) Click
“Set Working Directory”
9. Set the
stop volume by screwing the “Drive volume adjuster” anti-clockwise
1:10 mixing à
~200 uL (4 turns)
10. Purge
the system with NP-H2O
1) Ensure
that the drive rams are pushed fully down
2) Set the
drive valves to the “Load Position” (horizontal)
3) Fill up
the “C” and “F” reservoir syringes with NP-H2O
4) Fill the
drive syringes by pushing down the plungers of the reservoir syringes
5) Expel
any air bubbles by flushing backwards and forwards between the drive syringes
and the reservoir syringes several times
6) Fill the
drive syringes with NP-H2O until both drive syringe pistons contact
with drive ram
7) Set the
drive valves to the “Drive Position” (vertical)
8) Empty
any contents of the stop syringe by pulling the stop syringe piston upward at
the “Empty Position” of stop valve
9) Change
the stop valve from “Empty Position” to “Drive Position”
10) Push
the NP-H2O from the drive syringe, through the flow circuit and into
the stop syringe by pulling the drive ram upward
11) Repeat
1) ~ 10) steps 3~4 times

Figure 1. Sample handling unit (from
Applied Photophysics SX manual)

Figure 2. Valve position (from
Applied Photophysics SX manual)

Figure 3. Stop Valve position (from
Applied Photophysics SX manual)
Sample loading
11. Purge
the system with
12. Load
buffer in the drive syringe C (“Load Position”)
1) First,
fill the reservoir syringe with ~1mL of buffer and wash the drive syringe
2) Fill the
reservoir syringe with ~3mL of buffer and load in the drive syringe C
13. Load
denatured protein in the drive syringe F (“Load Position”)
1) Before
loading the protein, wash the drive syringe with 2.5M GuHCl (ß
same GuHCl concentration in which protein is denatured)
2)
Denatured protein should be filtered and degassed (ß
stirred for 10min under vacuum)
3) Load the
protein in the drive syringe F
4) Make
sure there is no air bubble
Program setting
14. Change
the signal mode to “Fluorescence”
15. Set the
excitation wavelength to 295
nm
16. Set the
detector voltages 400~500: has to be adjusted depending on protein signal. If
the signal exhibits saturation, low the voltage until full range of intensity
decrease is observed
17. Set the
Trigger mode to “External”
18. Click
“Seed” icon à type the file name
19. Set the
“time (s)”
20. Set the
number of data point (normally “1000” is recommended)
21. Set the
numbers of scan by clicking “Repeat”
Data collection
22.
Re-check both drive syringe pistons contact with drive ram
23. Change
the valve position to “Drive
Position” (vertical)
24. Prime
the flow circuit and cell contents by clicking “Drive” at least 5 times
25. Click
the “Acquire” to collect data

Figure 3. ProData SX window (from
Applied Photophysics SX manual)
To clean up the flow circuit and
cell contents after data collection
1. Purge
the system with 2.5M GuHCl
2. Purge
the system with > 10mL of NP-H2O
2. Run the
water-water scans (overlay of 5~6 scans)
3. Print
the spectra and record “Date”, “Voltage”, and “Your name” à
file in the SX20 log-in book
To turn off the instrument
1. Turn off
the “LAMP” à Record the lamp hour in the log-in
book
2.
Terminate “Pro-Data Viewer” and “Pro-Data SX” window
3. Turn off
the computer
4. Turn off
the “SYSTEM”
5. Close
the N2 gas
6. Release
holding pressure using the adjuster behind “sample handling unit”